The UDS-P LED by Woodpecker is a high-performance ultrasonic scaler engineered for precision, safety, and efficiency. It features a detachable, autoclavable LED handpiece, ideal for scaling and endodontic treatments, with advanced digital controls that deliver stable performance across a range of clinical applications.


Packaging

1 x Main unit

1 x Handpiece

1 x Foot switch

6 x Tip

1 x Adapter

1 x Torque wrench

1 x Operation manual

Direction to Use

Power Supply Requirements

  • Ensure the input voltage is within 220–240V~, 50/60Hz, 150mA.

  • Connect the main unit to a 24V~, 50/60Hz, 1.3A power source.


 Setup & Pre-Treatment Protocol

  1. Handpiece Installation:

    • Securely attach the handpiece to the unit.

    • Ensure it is detachable for autoclaving.

  2. Sterilization (Before Use):

    • Autoclave the following items:

      • Handpiece

      • Scaling tip

      • Torque wrench

      • Endo wrench

      • Endochuck

    • Follow recommended autoclaving methods (135°C, 0.22MPa).

  3. Tip Installation:

    • Use the torque wrench to screw the scaling tip onto the handpiece in the specified direction.


 Treatment Operation

  • Water Flow:

    • Ensure continuous water flow during operation to:

      • Cool the scaling tip

      • Maintain efficiency

  • Operating Mode:

    • Use the scaler in continuous mode.

    • Adjust power gradually based on treatment requirements.


 Post-Treatment Protocol

  1. Handpiece Care:

    • Unscrew and remove the scaling tip and endo chuck.

    • Sterilize the handpiece separately.

    • Allow it to cool naturally before reuse (to prevent scalding).

  2. Sterilization Guidelines:

    • Use sterile gauze or sterilization pouches for packaging.

    • Do not use:

      • Boiling water

      • Iodine

      • Alcohol

      • Chemical disinfectors


 Routine Maintenance

  • Handpiece Check:

    • Inspect regularly for damage or wear.

    • Lubricate “O” rings with dental-grade lubricant.

  • Connector & Cable:

    • Keep both the connector and cable socket dry and clean.

  • Daily Cleaning:

    • Wipe down the scaler before and after each use.

    • Maintain strict hygiene for optimal performance and infection control.

Warranty
  • 1-Year Warranty on all parts of the product

  •  5-Year Warranty on the PCB (Power Circuit Board) from the date of invoice

  • Physical damage is not covered under any circumstances